Specifications and Characteristics
2
N8480 Series Power Sensors Operating and Service Guide
31
Power Linearity
After zero and calibration at ambient environment conditions.
N O T E
The N8480 Series power sensors’ linearity is negligible except for the power range
specified in
.
Table 2-4
Power linearity
Sensor Model
Power Range
Linearity
(25°C ± 10 °C )
Linearity
(0 °C to 55 °C )
N8481A
N8482A
N8485A
N8487A
N8488A
N8486AR
N8486AQ
–1 dBm to +15 dBm
±0.52%
±0.80%
+15 dBm to +20 dBm
±0.80%
±1.90%
N8481B
N8482B
+29 dBm to +39 dBm
±0.52%
±0.80%
+39 dBm to +44 dBm
±1.66%
±2.75%
N8481H
N8482H
+17 dBm to +30 dBm
±0.77%
±1.05%
+30 dBm to +35 dBm
±2.84%
±3.93%