2
Specifications and Characteristics
40
N8480 Series Power Sensors Operating and Service Guide
Table 2-8
Calibration factor uncertainty at
25 ºC ± 3 ºC (continued)
Frequency
25 ºC ± 3 ºC
N8488A
10 MHz to 50 MHz
1.70
50 MHz to 100 MHz
1.60
100 MHz to 2 GHz
1.58
2 GHz to 12.4 GHz
1.75
12.4 GHz to 18 GHz
1.92
18 GHz to 26.5 GHz
2.43
16.5 GHz to 50 GHz
4.56
50 GHz to 67 GHz
5.30
67 GHz to 70 GHz
5.81
Table 2-9
Calibration factor uncertainty at
25 ºC ± 10 ºC
Frequency
25 ºC ± 10 ºC
N8481A
N8481B
N8481H
N8482A
N8482B
N8482H
100 kHz to 10 MHz
-
-
-
1.28
2.40
0.99
10 MHz to 30 MHz
1.47
1.98
1.39
1.03
1.84
0.86
30 MHz to 500 MHz
1.03
1.91
1.10
1.03
1.83
0.98
500 MHz to 1.2 GHz
0.97
1.91
1.15
1.08
1.94
1.00
1.2 GHz to 6 GHz
1.23
1.97
1.42
1.81
2.68
1.19
6 GHz to 14 GHz
1.85
2.69
3.87
-
-
-
14 GHz to 18 GHz
2.17
3.20
7.01
-
-
-