Keysight N5991 SATA Test Automation Software Platform User Guide
25
Using the Software
3
Once all the procedures are run, the N5991 configuration can be stored as
a single “.vfp” file using the
Save
button and recalled using
Load
button
without configuring the DUT again.
CAUTION
Before executing the calibration or test procedures, ensure that the SATA
Station Configuration has been conducted properly with all necessary
instruments such as the Infiniium oscilloscope set to “online”. All
calibrations can be run in offline mode, that is, without any instrument
connected. The offline mode is intended for product demonstrations with
simulated data. CALIBRATIONS RUN IN OFFLINE MODE DO NOT
GENERATE VALID CALIBRATION DATA.