![Keysight Technologies N5261A User'S And Service Manual Download Page 71](http://html1.mh-extra.com/html/keysight-technologies/n5261a/n5261a_users-and-service-manual_1961871071.webp)
Service Information
Over
c
urrent LEDs are On (
A
mber)
1.
Remove all millimeter module connections from the front panel. If the over
current LEDs are off, suspect the millimeter module or the DC Power Bias
cable. If the LEDs are on continue with
step 2
.
2.
Verify that the +12 Vdc on the front panel DC Power connectors is not
“Part Replacement and Location Information”
. If it is shorted to ground, replace the front panel DC Power
cable (N5652-60001) or the DC Power Board (N5261-60002). If it is not
shorted to ground continue to
step 3
.
3.
Move the DC Power board ribbon cable to the Source/Receiver Status LED
board, if the over current LEDs are still on replace the DC Power Board
(N5261-60002). If the LEDs are off replace the DC Power LED Board
(N5261-60005).
Front Panel Active and Port Status LED
Front Panel R and S Indicator LED check. If the LED indicators are not
operating, verify the ribbon cable connection and ensure the rear panel test set
interface cable is connected securely. Proceed to the next section for further
troubleshooting before replacing the front panel board.
1.
Connect the Test Set I/O cable (N4011-21002) from the analyzer to the
test set.
2.
Using the I/O command values, confirm the correct address and data
values are used, refer to
3.
Front Panel R and S indicator LED Check.
a.
Verify that at least three of the Controller Status LEDs, are on. Refer
b.
If none are on remove the Switch Driver board and recheck, if still no
indication, replace the Controller board.
c.
If the Controller Status LEDs and the front panel Active LED is on,
suspect the front panel LED board or the ribbon cable. Replace as
needed. You can move the ribbon cable connection to the DC power
board and verify if the problem is the Port Status LED board, or use
the ribbon cable from the DC power board to verify if the problem is
the ribbon cable.
Keysight N5261/62A User’s & Service Guide
6
5
Summary of Contents for N5261A
Page 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Page 6: ......
Page 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Page 54: ...48 Keysight N5261 62A User s Service Guide...
Page 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...