Service Information
2-
Adjust ALC Levels
Background
When a component is changed in an RF or LO signal path, it is often necessary
to adjust the ALC level for the path. See
8
for required
adjustments when a part is replaced.
PNA and Power Sensor Preparation
A power sensor and power meter must be prepared to make measurements at
8 GHz. This may include calibrating the power sensor with the meter and
entering a calibration factor for 8 GHz.
1.
Set the PNA to measure S12.
2.
Set for CW sweep at 8 GHz.
3.
Set the sweep time to 5 seconds.
4.
Set the power to -8 dBm.
ALC level for Port 1 (Procedure Q)
1.
Prepare the PNA and Power Sensor as described above.
2.
Connect PNA Port 2 to SRC 1 IN.
3.
Connect the power sensor to Port 1 RF OUT.
4.
Issue the Test Set I/O commands: 32,0 and 0,1 (ALC on and Port 1 RF
OUT).
5.
Identify the “Port 1 & 2 ALC Amp”, see
. Identity R7
and R3, see
6.
Adjust R7: turn full CW (may require up to 30 turns) then 5 turns CCW.
7.
Adjust R3: adjust for a reading on the power meter of 12.5 dB (± 0.1 dBm).
ALC level for Port 3 (Procedure R)
1.
Prepare the PNA and Power Sensor as described above.
2.
Connect PNA Port 2 to SRC 2 IN.
3.
Connect the power sensor to Port 3 RF OUT.
4.
Issue the Test Set I/O commands: 32,0 and 0,16 (ALC on and Port 3 RF
OUT).
5.
Identify the “Port 3 & 4
Mod/
Amp,” R7 and R3, see
6.
Adjust R7: turn full CW (may require up to 30 turns) then 5 turns CCW.
7.
Adjust R3: adjust for a reading on the power meter of 12.5 dB (± 0.1
dBm).
Keysight N5261/62A User’s & Service Guide
83
Summary of Contents for N5261A
Page 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Page 6: ......
Page 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Page 54: ...48 Keysight N5261 62A User s Service Guide...
Page 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...