![Keysight Technologies N5261A User'S And Service Manual Download Page 45](http://html1.mh-extra.com/html/keysight-technologies/n5261a/n5261a_users-and-service-manual_1961871045.webp)
Keysight N5261/62A User’s & Service Guide
39
Installation & System Operation
Configuring the Network Analyzer
to Millimeter Mode
This section will describe how to set up and operate the N5261/62A Millimeter
Head Controller with your network analyzer.
The N5261/62A Millimeter Head Controller is considered a “slave” instrument.
A network analyzer must be used to control the test set.
Before turning on this instrument, verify that the AC supply voltage is in the
specified range.
Typeface Key Conventions
The following key conventions are used throughout this document.
— [HARDKEYS] are labeled front panel keys.
— SOFTKEYS are indicated on the instrument display.
Before Beginning this Section
Determine the multiplier factors to use for RF IN and LO IN. The factors for
each head are indicated on the labels located on the top of each head (near
the RF and LO connectors).
The maximum frequency of RF signals from the PNA-X rear panel is 19 GHz.