Service Information
Table 2-
3
IF Path Measurement
From: TEST IF
Front Panel
To: N5261A Rear
Panel IF OUT
To: N5262A Rear
Panel IF OUT
Command
1
Max Path Loss for
N5262A (dB)
Port 1 TEST IF
A
A
16,8
3
Port 2 TEST IF
B
B
16,4
3
Port 3 TEST IF
n/a
C
16,2
3
Port 4 TEST IF
n/a
D
16,1
3
Port 1 REF IF
C
D
16,8
6.0
Port 1 REF IF
n/a
R
16,33
7.5
Port 2 REF IF
D
C
16,4
6.0
Port 2 REF IF
n/a
R
16,2
7.5
Port 3 REF IF
n/a
A
16,2
6.0
Port 3 REF IF
n/a
R
16,56
7.5
Port 4 REF IF
n/a
B
16,1
6.0
Port 4 REF IF
n/a
R
16,20
7.5