Performance Verification
5
N2797A User’s Guide
39
Test 1. DC Input Resistance
Procedure
1
Connect the DMM probes between the probe tip and ground at the tip of the probe.
2
Set up the DMM to measure resistance. The resistance should read 1 M
W
± 3%.
3
Record the resistance in