30
N2797A User’s Guide
2
Probe Calibration
Calibrating the Probe on InfiniiVision Oscilloscopes
Allow the oscilloscope and probe to warm up for 20 minutes before performing the probe
calibration.
Procedure
1
Connect the probe to one of the oscilloscope channels.
2
Connect the probe to the calibration terminals. On 3000 X-series oscilloscopes, connect
the probe’s signal tip to the front-panel
Demo 2 / Probe Comp
terminal and the ground tip to
the ground terminal.
Figure 12
Probe Calibration Terminals on 3000 X-Series Oscilloscope
3
Press the Channel on/off key to turn the channel on (if the channel is off).
4
In the Channel Menu, press the
Probe
softkey.
5
Press the
Calibrate Probe
softkey and follow the instructions on the display.
NOTE
When you connect the probe, the
Calibrate Probe
softkey in the Channel Probe Menu
becomes active.