Nominal Setup
Setup
DC level
SCPI command
76
[a]
-10.8
Output of 0 dB range
CAL:SET 76
77
+0.044 V
Output of -48 dB range (High DC range)
CAL:SET 77
78
[a]
-0.044 V
Output of -48 dB range (High DC range)
CAL:SET 78
79
+0.1 V
Output of -40 dB range (High DC range)
CAL:SET 79
80
[a]
-0.11 V
Output of -40 dB range (High DC range)
CAL:SET 80
81
+0.28 V
Output of -32 dB range (High DC range)
CAL:SET 81
82
[a]
-0.28 V
Output of -32 dB range (High DC range)
CAL:SET 82
83
+0.68 V
Output of -24 dB range (High DC range)
CAL:SET 83
84
[a]
-0.68 V
Output of -24 dB range (High DC range)
CAL:SET 84
[a] Constants are stored after completing this setup.
3. Enter the value measured using SCPI command.
4. To proceed:
a. If your calibration procedures require you to verify this adjustment, exit the calibration menu and perform
AC Amplitude (high-impedance) Verification
b. If you are making all of the adjustments and then verifying the instrument’s performance, continue with the
next procedure in this section.
44
Keysight EDU33210 Series Service Guide