Calibration Error
The following errors indicate failures that may occur during a calibration.
Error Code
Error Messages
Descriptions
701
Calibration error; security defeated by hardware jumper
If you short the calibration secure jumper (CAL ENABLE)
while turning the instrument on, this error indicates the
security password has been overwritten. See
for details.
702
Calibration error; calibration memory is secured
To perform calibration, unsecure the instrument. See
for details.
703
Calibration error; secure code provided was invalid
Specified security code was invalid.
706
Calibration error; value out of range
Value entered was outside valid range.
707
Calibration error; signal input is out of range
Occurs during the ADC Adjustment, setup 6, if the 1 V input
voltage is too high. May also occur during self-calibration
(setup 7). Run self-test to diagnose problem.
710
Self-calibration failed; Chan n, null DAC cal, invalid self cal
Self-calibration failed; Chan n, offset DAC cal with attenuator,
invalid self cal
Self-calibration failed; Chan n, offset DAC cal no attenuator,
invalid self cal
Error occurred while performing internal calibration of
specified DAC. Self-calibration exited without changing self-
calibration constants. Run self-test to diagnose problem.
711
Self-calibration failed; Chan n, null DAC cal gain too low (too
high), <meas_value>
Self-calibration failed; Chan n, offset DAC cal with attenuator
gain too low (too high), <meas_value>
Self-calibration failed; Chan n, offset DAC cal no attenuator
gain too low (too high), <meas_value>
Computed gain calibration factor for specified DAC was out
of limits. Self-calibration exited without changing self-
calibration constants. Run self-test to diagnose problem.
712
Self-calibration failed; Chan n, null DAC cal zero too low (too
high), <meas_value>
Self-calibration failed; Chan n, offset DAC cal with attenuator
zero too low (too high), <meas_value>
Self-calibration failed; Chan n, offset DAC cal no attenuator
zero too low (too high), <meas_value>
Self-calibration failed; Chan n, GND measurement out of
limits, <meas_value>
Computed zero calibration factor for specified DAC was out
of limits. Self-calibration exited without changing self-
calibration constants. Run self-test to diagnose problem.
715
Self-calibration failed; Chan n, null DAC cal, convergence
error sub attenuator value dB
Internal null DAC calibration failed to converge during
internal calibration. Self-calibration exited without changing
self-calibration constants. Run self-test to diagnose
problem.
720
Self-calibration failed; Chan n, offset DAC cal with attenuator,
convergence error
Self-calibration failed; Chan n, offset DAC cal no attenuator,
convergence error
Internal offset DAC calibration failed to converge internal
calibration. Self-calibration exited without changing self-
calibration constants. Run self-test to diagnose problem.
850
Calibration error; set up is invalid
Invalid calibration setup number selected.
Keysight EDU33210 Series Service Guide