AC amplitude (high-impedance) adjustment (channel 2)
The instrument stores a calibration constant for each high-impedance attenuator path. Each path's gain coefficient
is calculated using two measurements: one with the waveform DAC at + output and one with waveform DAC at –
output. The setups, therefore, must be performed in pairs.
1. Connect the multimeter to the channel output as shown below.
2. Use the multimeter to measure the DC voltage at the front panel connector for each setup in the following table.
Nominal Setup
Setup
DC level
SCPI command
57
+0.0028 V
Output of -72 dB range
CAL:SET 57
58
[a]
-0.0028 V
Output of -72 dB range
CAL:SET 58
59
+0.007 V
Output of -64 dB range
CAL:SET 59
60
[a]
-0.007 V
Output of -64 dB range
CAL:SET 60
61
+0.017 V
Output of -56 dB range
CAL:SET 61
62
[a]
-0.017 V
Output of -56 dB range
CAL:SET 62
63
+0.044 V
Output of -48 dB range
CAL:SET 63
64
[a]
-0.044 V
Output of -48 dB range
CAL:SET 64
65
+0.11 V
Output of -40 dB range
CAL:SET 65
66
[a]
-0.11 V
Output of -40 dB range
CAL:SET 66
67
+0.28 V
Output of -32 dB range
CAL:SET 67
68
[a]
-0.28 V
Output of -32 dB range
CAL:SET 68
69
+0.68 V
Output of -24 dB range
CAL:SET 69
70
[a]
-0.68 V
Output of -24 dB range
CAL:SET 70
71
+1.7 V
Output of -16 dB range
CAL:SET 71
72
[a]
-1.7 V
Output of -16 dB range
CAL:SET 72
73
+4.3 V
Output of -8 dB range
CAL:SET 73
74
[a]
-4.3 V
Output of -8 dB range
CAL:SET 74
75
+10.8
Output of 0 dB range
CAL:SET 75
Keysight EDU33210 Series Service Guide