![Keysight Technologies EDU33210 Series Service Manual Download Page 40](http://html1.mh-extra.com/html/keysight-technologies/edu33210-series/edu33210-series_service-manual_1962186040.webp)
8. Enter the value measured by multimeter (Step 4) using SCPI command.
9. Calculate the crossover point at 100 kHz by using below formula:
Crossover @100 kHz = Power meter converted value in Vrms (Step 7) – Multimeter measured value in Vrms
(Step 4)
10. Using the existing connection, configure the instrument and measure the output for each setup in the following
table.
Setup
Frequency
Amplitude
SCPI command
40
[a]
1 MHz
0.192 Vrms
CAL:SET 40
41
[a]
5 MHz
CAL:SET 41
42
[a]
10 MHz
CAL:SET 42
43
[a]
20 MHz
CAL:SET 43
44
[a]
25 MHz
CAL:SET 44
45
[a]
30 MHz
CAL:SET 45
[a] Constants are stored after completing this setup.
11. Convert the measured power level from dBm to Vrms.
12. Enter the value measured (calibration value) using SCPI command.
Calibration Value = Power meter converted value in Vrms – crossover @100 kHz in Vrms (Step 9)
13. To proceed:
a. If your calibration procedures require you to verify this adjustment, exit the calibration menu and perform
24 dB Range Flatness Verification
b. If you are making all of the adjustments and then verifying the instrument’s performance, verify the output
specifications of the instrument with the
40
Keysight EDU33210 Series Service Guide