AC amplitude (high-impedance) adjustment
The instrument stores a calibration constant for each high-impedance attenuator path. Each path's gain coefficient
is calculated using two measurements: one with the waveform DAC at + output and one with waveform DAC at –
output. The setups, therefore, must be performed in pairs.
1. Connect the multimeter to the channel output as shown below.
2. Use the multimeter to measure the DC voltage at the front panel connector for each setup in the following table.
Nominal Setup
Setup
DC level
SCPI command
10
+0.0028 V
Output of -72 dB range
CAL:SET 10
11
[a]
-0.0028 V
Output of -72 dB range
CAL:SET 11
12
+0.007 V
Output of -64 dB range
CAL:SET 12
13
[a]
-0.007 V
Output of -64 dB range
CAL:SET 13
14
+0.017 V
Output of -56 dB range
CAL:SET 14
15
[a]
-0.017 V
Output of -56 dB range
CAL:SET 15
16
+0.044 V
Output of -48 dB range
CAL:SET 16
17
[a]
-0.044 V
Output of -48 dB range
CAL:SET 17
18
+0.11 V
Output of -40 dB range
CAL:SET 18
19
[a]
-0.11 V
Output of -40 dB range
CAL:SET 19
20
+0.28 V
Output of -32 dB range
CAL:SET 20
21
[a]
-0.28 V
Output of -32 dB range
CAL:SET 21
22
+0.68 V
Output of -24 dB range
CAL:SET 22
23
[a]
-0.68 V
Output of -24 dB range
CAL:SET 23
24
+1.7 V
Output of -16 dB range
CAL:SET 24
25
[a]
-1.7 V
Output of -16 dB range
CAL:SET 25
26
+4.3 V
Output of -8 dB range
CAL:SET 26
27
[a]
-4.3 V
Output of -8 dB range
CAL:SET 27
28
+10.8
Output of 0 dB range
CAL:SET 28
Keysight EDU33210 Series Service Guide