DC offset voltage verification
This test checks the DC Offset Voltage on two attenuator ranges:
1. Set the multimeter to measure DCV. Connect the multimeter to the channel output as shown below.
2. Set the instrument to each output in the table below and measure the output voltage with the multimeter. Be
sure the output impedance is set to High–Z and the output is enabled.
Waveform Generator
Measurement
Output Setup
Function
Voltage
Nominal
Error
[a]
High Z
DC
0.0 V
0.0 VDC
±0.005 VDC
High Z
DC
500 mV
0.500 VDC
±0.010 VDC
High Z
DC
10.0 V
10.0 VDC
±0.105 VDC
[a] Based upon 1% of setting ±5 mVDC for High-Z.
3. Compare the measured value to the test limits shown in the table.
4. For EDU33212A only. Connect multimeter to channel 2 output and repeat steps 2 and 3.