Channel 1 Adjustments
Frequency (internal timebase) adjustment
AC amplitude (high-impedance) adjustment
-8 dB range flatness adjustment
-24 dB range flatness adjustment
Self-test
Run self-test to ensure that the instrument is in working order before beginning any additional adjustments. Be sure
to unlock the instrument and follow the requirements listed in Test Considerations before beginning any
adjustments.
1. Enter command CALibration:SeTup 1 to begin.
Setup
SCPI command
1
Performs the self-test. The main output is dis-
abled during test.
CAL:SET 1
2. If the instrument fails any self-test, press [System] > Help > Error View to record the error code and message,
and contact Keysight support if necessary.
A complete self-test (*TST?) takes approximately 12 seconds.
Keysight EDU33210 Series Service Guide
31