50
1000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
To compensate passive probes
You should compensate your passive probes to match their characteristics to the
oscilloscope’s channels. A poorly compensated probe can introduce measurement
errors.
1
Perform the procedure
“To perform an Auto Scale on the Probe Comp
2
Press the channel key to which the probe is connected (
[1]
,
[2]
).
3
In the Channel Menu, press
Probe
.
4
In the Channel Probe Menu, press
Probe Check
; then, follow the instructions
on-screen.
If necessary, use a nonmetallic tool (supplied with the probe) to adjust the
trimmer capacitor on the probe for the flattest pulse possible.
On the some probes, the trimmer capacitor is the yellow adjustment on the
probe tip. On other probes, the trimmer capacitor is located on the probe BNC
connector.
Figure 7
Example pulses
5
Connect probes to all other oscilloscope channels (channel 2 of a 2-channel
oscilloscope).
6
Repeat the procedure for each channel.
The process of compensating the probes serves as a basic test to verify that the
oscilloscope is functional.
comp.cdr
Perfectly compensated
Over compensated
Under compensated
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...