Troubleshooting
4
1000 X-Series Oscilloscopes Service Guide
49
To perform an Auto Scale on the Probe Comp signal
1
Press the
[Default Setup]
key. The oscilloscope is now configured to its default
settings.
2
Connect an oscilloscope probe from channel 1 to the
Probe Comp
signal terminal
on the front panel.
3
Connect the probe’s ground lead to the ground terminal that is next to the
Demo
2
(Probe Comp) terminal.
4
Press
[AutoScale]
.
5
You should see a waveform on the oscilloscope’s display similar to this:
If you see the waveform, but the square wave is not shaped correctly as shown
above, perform the procedure
“To compensate passive probes"
If you do not see the waveform, ensure your power source is adequate, the
oscilloscope is properly powered-on, and the probe is connected securely to
the front-panel analog channel input BNC and to the Demo 2 (Probe Comp)
terminal.