34
1000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Figure 4
Setting Up Equipment for Internal Trigger Sensitivity Test
4
Set the output frequency of the signal generator to 10 MHz.
5
Perform these steps to test at the 10 MHz setting:
a
Set the signal generator amplitude to about 20 mV
pp
.
b
Press the
[AutoScale]
key.
c
Set the time base to 10 ns/div.
d
Set channel 1 to 10 mV/div.
e
Decrease the amplitude from the signal generator until 0.6 vertical divisions
of the signal (about 6 mV
pp
) is displayed.
50 Ohm Feedthrough
Signal
Generator
Oscilloscope
N Cable
N to BNC Adapter