Testing Performance
2
1000 X-Series Oscilloscopes Service Guide
39
Keysight 1000 X-Series Oscilloscopes Performance Test Record
Serial No. ______________________________________
Test by _____________________________
Test Interval ____________________________________
Work Order No. ______________________
Recommended Next Testing ________________________
Temperature ____________
DC Vertical Gain Accuracy
Range
Power Supply Setting
Test Limits
Channel 1 Channel 2
10 V/Div
70 V
67.6 V to 72.4 V
________
________
5 V/Div
35 V
33.8 V to 36.2 V
________
________
2 V/Div
14 V
13.52 V to 14.48 V
________
________
1 V/Div
7 V
6.76 V to 7.24 V
________
________
500 mV/Div
3.5 V
3.38 V to 3.62 V
________
________
200 mV/Div
1.4 V
1.352 V to 1.448 V
________
________
100 mV/Div
700 mV
676 mV to 724 mV
________
________
50 mV/Div
350 mV
338 mV to 362 mV
________
________
20 mV/Div
140 mV
135.2 mV to 144.8 mV
________
________
10 mV/Div
70 mV
67.6 mV to 72.4 mV
________
________
5 mV/Div
35 mV
33.4 mV to 36.6 mV
________
________
2 mV/Div
14 mV
13.36 mV to 14.64 mV
________
________
1 mV/Div
7 mV
6.68 mV to 7.32 mV
________
________
0.5 mV/Div
3.5 mV
3.18 mV to 3.82 mV
________
________
Band wid th (-3 dB)
Model
Test Limits
Channel 1 Channel 2
100 MHz
-3 dB at 100 MHz
________
________
70 MHz
-3 dB at 70 MHz
________
________
50 MHz
-3 dB at 50 MHz
________
________
Time Base Accuracy Limits
Calculated
time base
accuracy
limit (ppm)
Measured
time base
error (ppm)
Pass/Fail
Time Base Accuracy Limit: 50 ppm ±5 ppm per year
(aging)
_________ ________ ________
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...