Testing Performance
Test External Trigger Sensitivity
This test applies to all models.
Verify the external trigger sensitivity at these settings:
•
10 MHz, 250 mV
pp
in 8 V range
•
10 MHz, 50 mV
pp
in 1.6 V range (DSOX models)
•
Max BW of oscilloscope, 500 mV
pp
in 8 V range
•
Max BW of oscilloscope, 100 mV
pp
in 1.6 V range (DSOX models)
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the power splitter input.
b
Connect one output of the power splitter to the Ext Trig input through a 50
Ω
feedthrough termination.
c
Connect the power sensor to the other output of the power splitter.
Table 11
External Trigger Sensitivity Specifications
Oscilloscope
Models
Range
Setting
External Trigger Sensitivity Specification Band wid th
<= 10 MHz
Maximum Band wid th
50 MHz
8 V
250 mVpp
500 mVpp
70 MHz
8 V
250 mVpp
500 mVpp
1.6 V
50 mVpp
100 mVpp
100 MHz
8 V
250 mVpp
500 mVpp
1.6 V
50 mVpp
100 mVpp
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...