Testing Performance
2
1000 X-Series Oscilloscopes Service Guide
29
11
Set the oscilloscope sweep speed according to the following:
•
100 MHz Models: 5 ns/div
•
70 MHz Models: 5 ns/div
•
50 MHz Models: 5 ns/div
12
Note the oscilloscope AC RMS - Cyc(1) reading at the bottom of the screen.
13
Note the reading on the power meter and covert to Vrms using the expression:
14
Calculate the response using the expression:
Example
If:
Pmeas
1_MHz
= 892 uW
Std Dev(n)
1MHz
= 210.4 mV
Pmeas
max_freq
= 687 uW
Std Dev(n)
max freq
= 161.6 mV
Then after converting the values from the power meter to Vrms:
15
The result from step 14 should be b3.0 dB and -3.0 dB. Record the
result in the Performance Test Record (see
16
Move the power splitter from the channel 1 to the channel 2 input.
17
Turn off the current channel and turn on channel 2 using the channel keys.
18
Set the trigger source to channel 2.
19
Repeat steps 3 through 15 for the remaining channels, setting the parameters
of the channel being tested where appropriate.
Vin
maxfreq
Pmeas
maxfreq
50
Ω
×
=
response(dB) = 20 log
10
MHz
1
M Hz
1
freq
max
freq
max
Vin
/
Vout
Vin
/
Vout
response(dB) = 20 log
10
mV
211.2
mV/
210.4
mV
185.3
/
mV
161.6
= -1.16 dB
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...