Testing Performance
2
1000 X-Series Oscilloscopes Service Guide
35
The trigger is stable when the displayed waveform is stable. If the trigger is
not stable, try adjusting the trigger level. If adjusting the trigger level makes
the trigger stable, the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
f
Record the result as Pass or Fail in the Performance Test Record (see
).
g
Repeat this step for the remaining oscilloscope channels.
6
Set the output frequency of the signal generator to the maximum bandwidth of
the oscilloscope:
•
100 MHz models: 100 MHz
•
70 MHz models: 70 MHz
•
50 MHz models: 50 MHz
7
Perform these steps to test at the maximum bandwidth setting:
a
Set the signal generator amplitude to about 20 mV
pp
.
b
Press the
[AutoScale]
key.
c
Set the time base to 10 ns/div.
d
Set channel 1 to 10 mV/div.
e
Decrease the amplitude from the signal generator as described in the
following table according to your oscilloscope’s maximum bandwidth:
The trigger is stable when the displayed waveform is stable. If the trigger is
not stable, try adjusting the trigger level. If adjusting the trigger level makes
the trigger stable, the test still passes. If adjusting the trigger does not help,
see the “Troubleshooting” chapter. Then return here.
f
Record the result as Pass or Fail in the Performance Test Record (see
).
g
Repeat this step for the remaining oscilloscope channels.
Oscilloscope’s maximum
band wid th:
Decrease amplitude until these
vertical divisions are displayed:
Which is about this V:
50 MHz, 70 MHz
0.9 vertical divisions
9 mVpp
100 MHz
1.2 vertical divisions
12 mVpp
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...