22
1000 X-Series Oscilloscopes Service Guide
2
Testing Performance
3
Set up the oscilloscope.
a
Adjust the horizontal scale to
200.0 us/div
.
b
Set the Volts/Div setting to the value in the first line and column of
c
Adjust the channel’s vertical position knob to place the baseline (reference
level) at 0.5 major division from the bottom of the display.
Table 5
Settings Used to Verify DC Vertical Gain Accuracy
Volts/Div Setting
Oscilloscope Calibrator
Setting
Test Limits
10 V/Div
70 V
67.6 V
to
72.4 V
5 V/Div
35 V
33.8 V
to
36.2 V
2 V/Div
14 V
13.52 V
to
14.48 V
1 V/Div
7 V
6.76 V
to
7.24 V
500 mV/Div
3.5 V
3.38 V
to
3.62 V
200 mV/Div
1.4 V
1.352 V
to
1.448 V
100 mV/Div
700 mV
676 mV
to
724 mV
50 mV/Div
350 mV
338 mV
to
362 mV
20 mV/Div
140 mV
135.2 mV
to
144.8 mV
10 mV/Div
70 mV
67.6 mV
to
72.4 mV
5 mV/Div
1
35 mV
33.4 mV
to
36.6 mV
2 mV/Div
1
14 mV
13.36 mV
to
14.64 mV
1 mV/Div
1
7 mV
6.68 mV
to
7.32 mV
0.5 mV/Div
1, 2
3.5 mV
3.18 mV
to
3.82 mV
1
A blocking capacitor is required at this range to reduce noise. See
2
Full scale is defined as 8 mV on the 500 uV/div range. Full scale on all other ranges is defined as 8
divisions times the V/div setting.
Summary of Contents for DSOX1102A
Page 1: ...Keysight InfiniiVision 1000 X Series Oscilloscopes Service Guide ...
Page 7: ...1000 X Series Oscilloscopes Service Guide 7 Contents Index ...
Page 8: ...8 1000 X Series Oscilloscopes Service Guide Contents ...
Page 10: ...10 1000 X Series Oscilloscopes Service Guide Figures ...
Page 12: ...12 1000 X Series Oscilloscopes Service Guide Tables ...
Page 52: ...52 1000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 54: ...54 1000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 60: ...60 1000 X Series Oscilloscopes Service Guide Index ...