Table 2-5
Electrical Specifications for 1.0 mm 50
Ω
Devices
Device
Frequency Parameter
Specifications
Male
Female
Loads
DC to 2 GHz
2 GHz to 18 GHz
18 GHz to 40 GHz
40 GHz to 50 GHz
Return Loss
30 dB
30 dB
26 dB
24 dB
30 dB
30 dB
26 dB
24 dB
Opens
DC to 2 GHz
2 GHz to 18 GHz
18 GHz to 50 GHz
Deviation from
Nominal Phase
±
1.0 degrees
±
1.5 degrees
±
2.5 degrees
±
1.0 degrees
±
3.0 degrees
±
4.0 degrees
Short 3
DC to 2 GHz
2 GHz to 18 GHz
18 GHz to 50 GHz
50 GHz to 110 GHz
Deviation from
Nominal Phase
±
0.8 degrees
±
1.2 degrees
±
1.5 degrees
±
3.0 degrees
±
1.0 degrees
±
2.0 degrees
±
2.5 degrees
±
5.0 degrees
Short1
50 GHz to 110 GHz
Deviation from
Nominal Phase
±
2.5 degrees
±
4.0 degrees
Short 2
75 GHz to 110 GHz
Deviation from
Nominal Phase
±
2.5 degrees
±
4.0 degrees
Short 4
50 GHz to 75 GHz
Deviation from
Nominal Phase
±
2 degrees
±
3.5 degrees
Device
Frequency Parameter
Specifications
Lossy Delay Line
DC to110 GHz
Return Loss
18 dB
Adapters
DC to 20 GHz
20 GHz to 50 GHz
50 GHz to 75 GHz
75 GHz to 110 GHz
Return Loss
24 dB
20 dB
18 dB
14 dB
Verification Match
Thru (adapter)
DC to 20 GHz
20 GHz to 50 GHz
50 GHz to 75GHz
75 GHz to 110 GHz
Return Loss
24 dB
20 dB
18 dB
14 dB
Verification
Mismatch Thru
(adapter)
DC to 110 GHz
Return Loss
6 dB @ ~ 22.6 GHz
intervals
Summary of Contents for 85059A
Page 2: ...Keysight Technologies 85059A 1 0 mm Calibration and Verification Kit...
Page 10: ...Contents 85059A TOC iv...
Page 30: ...3 2 Keysight Technologies 85059A 1 0 mm Calibration and Verification Kit Making Connections...
Page 62: ...8 2 Keysight Technologies 85059A 1 0 mm Calibration and Verification Kit Standard Definitions...