85059A 1.0 mm Calibration and Verification Kit
1-3
General Information
Achieving Specified Frequency Performance
Achieving Specified Frequency Performance
The standards in this calibration and verification kit allow you to perform
simple 1- or 2-port calibrations, as well as TRM (thru-reflect-match)
calibrations, and to verify the performance of your PNA system.
NOTE
Above 50 GHz, offset shorts are substituted for the opens and loads.
NOTE
For best results, before beginning calibration refer to
Calibration Standards
Table 1-1
Calibration Techniques and Standards
Frequency
Range
Calibration
Technique
“Open”
Standard
(Reflection)
“Short”
Standard
(Reflection)
“Load”
Standard
(Reflection)
“Thru”
Standard
(not used for
1-port cal)
“Isolation”
Standard
1
DC - 50 GHz
SOLT
2
Open
Short 3
50 GHz Load
Thru
50 GHz Load
50 - 75 GHz
Offset Shorts
Short 3
Short 1
Short 4
Thru
Load BB
3
75 - 110 GHz
Offset Shorts
Short 3
Short 1
Short 2
Thru
Load BB
3
To Combine Frequency Ranges
4
DC - 75GHz
SOLT
2
and
Offset Shorts
Open
Short 3
Short 3
Short 1
50 GHz Load
Short 4
Thru
Load BB
3
50 - 110 GHz
Offset Shorts
Short 3
Short 1
Short 4
Short 2
Thru
Load BB
3
DC - 110 GHz
SOLT
2
and
Offset Shorts
Open
Short 3
Short 3
Short 1
50 GHz Load
Short 4
Short 2
Thru
Load BB
3
1. For best measurement results, the isolation standard should be the equivalent impedance of the device under test.
2. SOLT = Short, Open, Load and Thru standards
3. The broadband load is a combination of a lossy delay line plus a 50 GHz load (see
4. Combinations of the calibration methods shown in the upper half of
are used to calibrate over the frequency ranges shown
in the lower half of the table.