85059A 1.0 mm Calibration and Verification Kit
5-1
5
Performance Verification
System Verification
After installation of the system is complete, a performance verification is
necessary to assure proper system operation. The initial verification is included
with the installation of your system, but it is necessary to perform system
verification at regular intervals. Included with this kit are a matched thru
adapter and a mismatched thru adapter for use in system verification. For more
information on system verification see the
analyzer’s
User’s Guide
or
Help System.
Calibration Kit
Verification
The performance of your calibration and verification kit can only be verified by
returning the kit to Keysight Technologies for recertification. The equipment
and calibration standards required to verify the specification limits of the
devices inside this kit have been specially manufactured, and are not
commercially available. Keysight recognizes its responsibility to provide you
with procedures to reconfirm the published specifications of any product
offered. That commitment applies equally to the 85059A 1.0 mm calibration and
verification kit.
What Recertification
Provides
The following will be provided with a recertified kit:
•
New calibration sticker affixed to the case
•
Certificate of Calibration
•
A calibration report for each device in this kit listing measured values,
specifications, and uncertainties.
NOTE
A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
For more information, contact the nearest Keysight Technologies office (sales
and service offices are listed in the front of this manual).
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The
actual need for recertification depends on the use of your kit. After reviewing
the results of the initial recertification, you may want to establish a different
recertification interval that reflects the usage and wear of your kit.
Summary of Contents for 85059A
Page 2: ...Keysight Technologies 85059A 1 0 mm Calibration and Verification Kit...
Page 10: ...Contents 85059A TOC iv...
Page 30: ...3 2 Keysight Technologies 85059A 1 0 mm Calibration and Verification Kit Making Connections...
Page 62: ...8 2 Keysight Technologies 85059A 1 0 mm Calibration and Verification Kit Standard Definitions...