(3) Calibrating the LE-200C • LH-200C • LZ-200C
The LE-200C • LH-200C • LZ-200C can store up to 4 sets each of
calibration parameters for both the electromagnetic and high-
frequency measurement formats.
Decide which application number you wish to store a particular set
of calibration parameters in before performing the calibration
procedure.
In order to obtain stable precision, we recommend that you
perform 10 to 20 practice measurements to calibrate the unit after
turning its power on.
1
The calibration Parameters application number (APPL.
No.) is specified. Allowable application numbers are
from 1 to 4.
【
Operation Example
】
: Set the Application Number to 4. Use the
LHP-20C probe when making this setting.
* FUNCTION KEY
* APPL. No=(1-4)?
(Number displayed is the previous
application number)
* APPL.No=4
* NF
(Input complete)
*** APPL.No=4 ***
(Printout)
1
2
3
4
14