This function mode is used to prevent incorrect measurement data
from being used in statistical calculations.
Please note that statistical calculation results already obtained for pre-
vious blocks cannot be changed.
【
Operation Example 1
】
: To erase data immediately after making an
incorrect measurement.
Because the Delete operation is performed immediately after making
the measurement "N = 004, B = 01, 143
µ
m", all the data for measure-
ment number N = 0004 is deleted. Therefore the next measurement
begins from measurement number N = 0004.
24
LIMIT (Upper and lower limits)
This function mode is used to set upper and lower limits when checking to
make sure that measured coating thicknesses fall within a specified range.
An electronic beep will sound and an exclamation point (!) will be printed
when a measurement value falls outside the specified range.
【
Operation Example
】
: Set upper limit at 100
µ
m and lower limit at 50
µ
m
Use the following procedure to erase upper and lower limit setting
values from memory.
➡
➡
➡
➡
➡
Key Operations
Display
# FUNCTION KEY
# =?
m
# =?
m
# =100?
m
# =?
m
# =?
m
# =50?
m
**U.LIMIT=100 m**
**L.LIMIT= 50 m**
Printout example
➞
Key Operations
Display
# FUNCTION KEY
# DELETE N=?
# NF(FE)
Printout example
➞