I — DESCRIPTION.

The BOP 1KW-MG/ME Series hereafter referred to
as BOP, are true 4-quadrant programmable voltage
and current power supplies, meaning they are capa-
ble of both sourcing and sinking power (see Figure
61). These bipolar power supplies pass smoothly
through zero without switching to provide true ± volt-
age and ± current. These BOP power supplies use
switch mode technology for low dissipation. A bi-
directional, isolating, a-c input power factor correct-
ing (PFC) circuit recuperates energy sinked from an
active load and sends it back into the line to maintain
low dissipation.

These BOP power supplies are controlled digitally
from a menu-driven front-panel keypad or one of the
remote digital interfaces (GPIB or RS 232 on MG

models, LAN or RS 232 on ME models) to set volt-
age and current and the four protection limits (+volt-
age, –voltage, +current and –current.) A front panel
rotary adjuster allows real-time adjustment of the
output. A large LCD displays the mode of operation,
the settings, and the actual output voltage and cur-
rent. Additionally, these BOP models can be
remotely controlled by an analog ±10V input for the
main channel (voltage or current), and a +1 to +10V
input for the limit channels. 

BOP models are suitable for driving inductive loads
such as large magnets or motors, and for exercising
batteries. They are also suitable for characterizing
solar cell arrays, and powering many electrochemi-
cal reactions.

II — UNPACKING.

This instrument has been thoroughly inspected and
tested prior to packing and is ready for operation.
After careful unpacking, inspect for shipping damage
before attempting to operate. Perform the “Installa-

tion.” on page 5. If any indication of damage is
found, file an immediate claim with the responsible
transport service.

Description.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
Equipment Supplied.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3
Safety.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3
Preliminary Operational Check.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3
Accessories.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
Installation.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5

Input Connections.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
Load Connections.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
Local Sensing (Factory Default). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
Remote Sensing Select.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
Analog I/O Connections.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
Trigger Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
GPIB Connections (MG SUFFIX).   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
LAN Connections (ME SUFFIX ONLY).  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
FINDING KEPCO POWER SUPPLIES ON THE LAN.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
RS 232 Connections.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7

Operation.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9

Turning The Power Supply On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9
Accessing the Menus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9
How to Modify a Parameter.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  10
Adjusting LCD Brightness, contrast and Background  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  11
Enabling/Disabling Audible Beeps   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  11
Setting Voltage or Current Mode   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  11
Programming Voltage or Current and Associated Protect Limits  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  11
Software Limits.   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  12

Changing Main Channel Software limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  12

Determining How the Unit Responds when Output is OFF (Load Type)   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  12
Configure Load Type.  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  14
Enabling/disabling Output Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  14

Additional Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  14
BOP 1KW Outline Dimensions   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  15

Summary of Contents for BOP 10-100MG/ME

Page 1: ...units the GPIB interface is replaced with a LAN interface Unless otherwise noted this guide applies to both models ACCESSING MANUALS First determine your Firmware Version see below then download the B...

Page 2: ...or current and a 1 to 10V input for the limit channels BOP models are suitable for driving inductive loads such as large magnets or motors and for exercising batteries They are also suitable for chara...

Page 3: ...el 1 d c Output Range Closed Loop Gain Output Impedance Voltage 2 V d c Current A d c Voltage Channel GV V V Current Channel GI A V Voltage Mode Series R L Current Mode Parallel R C Rd c mOhms L H Rd...

Page 4: ...4 228 1692 REV 8 090418 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 1 BOP OUTPUT CHARACTERISTICS...

Page 5: ...n to IEEE 1118 BITBUS port 142 0485 Kepco KMDLA 5P Kycon Inc IEEE 488 Cable 1 meter long For MG models only Connects BOP power supply to GPIB bus SNC 488 1 IEEE 488 Cable 2 meter long For MG models on...

Page 6: ...operation voltage or current output volt age or current and establish positive and negative voltage and current limits An output analog corre sponding to output current is also provided Refer to Oper...

Page 7: ...LS AND INDICATORS RS 232 CONNECTIONS Connect the BOP 1KW to a modem using a Null Modem patch cable at the RS 232 port located on the rear panel See Fig ure 3 A Null Modem cable is not required for old...

Page 8: ...1355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 3 BOP 1KW SERIES REAR PANEL VIEW LINKS INSTALLED FOR LOCAL SENSING FIGURE 4 LOAD CONNECTION A LOCAL...

Page 9: ...ng an item on the list and pressing the View Modify function key Menus and submenus will display a list of parameters with the top one high lighted The function key assignments can vary but generally...

Page 10: ...g and output Calibration Voltage Current and External and Controls calibration Shows Firmware versions for Display Serial GPIB LAN and analog Power up Settings Mode Voltage Cur rent External Main chan...

Page 11: ...channels one to set output voltage or current and one to set the corre sponding protection limit The main channel is deter mined by the MODE key in local mode which alternately selects either Voltage...

Page 12: ...ral Setup menu then highlight Max Min Set tings 2 Press to enter the Max Min Settings submenu Figure 6 If a Password is required see Opera tor s Manual for instructions 3 Highlight the voltage or curr...

Page 13: ...afety fea tures constant current type active electronic loads must be adjusted to zero and the BOP front panel LCD must read 0V minimum cur rent before handling the power supply to load connections Re...

Page 14: ...flow during the OFF state Table 5 indicates how the power supply responds to a command to go from Output ON to OFF CONFIGURE LOAD TYPE To configure press from the power up screen then highlight Load T...

Page 15: ...copower com email hq kepcopower com FIGURE 1 BOP 1KW OUTLINE DIMENSIONS SHEET 1 OF 2 20 000 508 00 21 439 544 54 22 800 579 11 21 564 547 73 22 000 558 79 23 863 606 12 18 235 463 16 18 985 482 21 1 4...

Page 16: ...FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 1 BOP 1KW OUTLINE DIMENSIONS SHEET 2 OF 2 SLIDES TRAVEL DISTANCE 23 000 584 2 22 000 5...

Reviews: