2.3 Rinse the electrode with distilled water and with a small amount of the next calibration solution. 
2.4 Dip the electrode into the 12.88 mS/cm buffer solutions Stir the tester gently and make sure the electrode is  
      completely submerged into the solution. The value of the calibration solution should appear. 
 
 
 
 
 
 
 
 
2.5 Press ENTER to confirm. The displayed value should begin flashing. Consult step 1.6 if the displayed value does  
      not match the value of the calibration solution. Wait for the measured value to stabilize.  
2.6 Press ENTER to confirm again to start the calibration. Wait for the reading to 

stabilize until the display shows ‘END’  

      and completes the calibration process. 
 

Three points Calibration 
 

3.1 Make sure to select three points calibration in the settings 
3.2 Do steps 1.2 to 1.5 and then steps 2.3 to 2.6 
3.3 When the second calibration point is completed, 

the display will show ‘CAL3” as the tester prompts you 

      to continue with the third point of calibration. 
 
 
 
 
 
 
 
3.4 Rinse the electrode with distilled water and small amount of the buffer solution. 
3.5 For the third point of calibration 111.8 mS/cm should be used. 
3.6 Press ENTER to confirm. The value displayed should be similar to the value of the used buffer solution.  
3.7 Follow step 1.6 to modify the value as needed. 
3.8 PRESS ENTER to confirm. Wait for the measur

ement to stabilized until the display shows ‘END’  

     and completes the calibration process. 
 

SWITCHING BETWEEN MEASUREMENT MODES 
 

1. Press MEAS to initiate measurement mode. The default measure is for salinity. To change to another  
    measurement mode, press ENTER to switch to conductivity mode.  
 

 
 
 
 
 
 
 

 
 
 

Summary of Contents for KS-SAL-20

Page 1: ...RTING THE BATTERIES 1 Your Kensington Pocket KS SAL 20 pocket salinity tester uses two AAA batteries 2 Remove the button cap found on the bottom of the tester as shown in the above image 3 Place the battery into its designated slot Take note of the proper position of the battery to ensure the polarities placed correctly 4 Replace the battery cap KEYPAD INDEX Key Function 1 Serves as the power butt...

Page 2: ... mode SETUP Indicates tester is in SETUP mode HOLD Indicates the HOLD function is active and the current displayed value is on hold ATC Indicates temperature compensation is enabled SETUP MENU The KS SAL 20 contains integrated setup menu that allows you to customize your tester s features to fit your measurement requirements Parameter Description Options Description CAL Calibration Points 1 Defaul...

Page 3: ...the ENTER key to confirm 3 The tester will now automatically display the temperature unit selection mode The display should show C UNIT 6 Press CAL to the select the chosen temperature unit C or F Press ENTER to confirm 7 The next display will show the temperature calibration mode The display should show C CAL 8 If you do not wish to conduct temperature calibration press ENTER to proceed with the ...

Page 4: ...how that auto off function is enabled 13 Press the CAL key to disable this feature and press ENTER to confirm 14 The final setting is for the Reset function The display should show NO RST 15 Press the CAL key to enable or disable the function and press ENTER to confirm Remember that enabling this functions will restore the tester to factory default setting and erase all selected parameters 16 The ...

Page 5: ... calibration point 1 4 Dip the conductivity electrode into the prepared buffer solution 1413 µS cm The end of the sensor must be completely submerged Stir the tester 1 5 Press the ENTER to confirm The tester should recognize the value of the used solution and the displayed value should begin to flash 1 6 If the displayed value is incorrect press the CAL button to set each digit and press ENTER to ...

Page 6: ... the display shows END and completes the calibration process Three points Calibration 3 1 Make sure to select three points calibration in the settings 3 2 Do steps 1 2 to 1 5 and then steps 2 3 to 2 6 3 3 When the second calibration point is completed the display will show CAL3 as the tester prompts you to continue with the third point of calibration 3 4 Rinse the electrode with distilled water an...

Page 7: ...D key again to disable the hold function ELECTRODE SENSOR MAINTENANCE 1 It is recommended to keep the sensitive membrane of the sensor wet This is accomplished by periodically dipping the sensor in distilled water or by using an electrode storage solution when the tester is not in use 2 If the instrument will be kept in storage for a long time place the sensor in an electrode storage solution 3 Ma...

Page 8: ...S cm 84µS cm 1413µS cm 12 88mS cm Conductivity Range 100 2000µS cm 2 00 20 00 20 00 200 0mS cm Accuracy 1 F S Resolution 0 01 0 1 1 Calibration points 1 to 3 points Calibration solutions 10µS cm 84µS cm 1413µS cm 12 88mS cm Temperature Range 0 60 C 32 140 F Accuracy 1 C Resolution 0 1 C 0 1 F Calibration point 1 Calibration range Measured value 10 C Others Temperature compensation 0 60 C 32 140 F ...

Reviews: