9. The next setting is for the HOLD function. The display should show YES/HOLD indicating the auto hold function is
enabled.
10. If the automatic hold function is enabled, the tester will automatically freeze the stable measured value. If this
function is disabled, the user will have to press the HOLD key to enable the function when needed.
11. Press CAL key to enable or disable the HOLD function and press ENTER to confirm.
12. The next display should show
‘YES/OFF’ to show that auto off function is enabled.
13. Press the CAL key to disable this feature and press ENTER to confirm.
14. The final setting is for the Reset function. The display should show NO/RST
15. Press the CAL key to enable or disable the function and press ENTER to confirm. Remember that enabling this
functions will restore the tester to factory default setting and erase all selected parameters.
16. The tester will automatically go to measurement mode.
CONDUCTIVITY CALIBRATION
It is recommended that the KS-SAL-20 should be calibrated regularly to ensure accurate measurements. The KS-SAL-
20 allows up to 3 points calibration and it is recommended to use 3 points calibration before measurement.
The tester is programmed to automatically detect the needed standard solutions for calibration and will prompt the user
during the calibration process.