background image

 

 

 

User Manual 

KS-SAL-20 |Pocket  

Salinity Tester 

 

Kensington Scientific 

22a Cavan Rd, Dry Creek 
South Australia, 5094 
[email protected] 
www.kensingtonscientific.com.au

 

 
2. Press ENTER again to return to salinity mode.  
 

 
 
 
 
 
 
 
MAKING MEASUREMENTS 
 

1. Press the MEAS/HOLD button to turn on the meter. Press ENTER to select the needed measurement mode. 
2. Remove the protective cap of the salinity sensor  
3. Rinse the meter with distilled water. 
4. Dip the instrument into the tested sample or solution and stir gently. 
5. Wait for the measurement to stabilize and record the measurement. 
 

HOLD FUNCTION 
 

1. Pressing the HOLD key while in measurement mode will activate the hold function. The HOLD indicator will appear  
    on the display and further measurement readings will not be possible. 
2. Press the HOLD key again to disable the hold function. 
 
 

 
 

 
 

 
 
 
ELECTRODE SENSOR MAINTENANCE 

 
1. It is recommended to keep the sensitive membrane of the sensor wet. This is accomplished by periodically dipping  
    the sensor in distilled water or by using an electrode storage solution when the tester is not in use. 
2. If the instrument will be kept in storage for a long time, place the sensor in an electrode storage solution. 
3. Make sure to wash the sensor with distilled water before and after use. 
4. Do not reuse prepared buffer solutions.  
  

ELECTRODE REPLACEMENT 
 

One of the advantages of using the Kensington KS-EC-10M-E is that the sensor can be replaced instead of buying a 
new tester when the sensor fails during calibration. Removing and installing the sensor is easy and convenient and 
also increases the utility of the instrument as additional sensors can be purchased for different applications. To replace 
the sensor: 
 
1. Twist the electrode collar counter-clockwise and pull the electrode away from the instrument 

Summary of Contents for KS-SAL-20

Page 1: ...RTING THE BATTERIES 1 Your Kensington Pocket KS SAL 20 pocket salinity tester uses two AAA batteries 2 Remove the button cap found on the bottom of the tester as shown in the above image 3 Place the battery into its designated slot Take note of the proper position of the battery to ensure the polarities placed correctly 4 Replace the battery cap KEYPAD INDEX Key Function 1 Serves as the power butt...

Page 2: ... mode SETUP Indicates tester is in SETUP mode HOLD Indicates the HOLD function is active and the current displayed value is on hold ATC Indicates temperature compensation is enabled SETUP MENU The KS SAL 20 contains integrated setup menu that allows you to customize your tester s features to fit your measurement requirements Parameter Description Options Description CAL Calibration Points 1 Defaul...

Page 3: ...the ENTER key to confirm 3 The tester will now automatically display the temperature unit selection mode The display should show C UNIT 6 Press CAL to the select the chosen temperature unit C or F Press ENTER to confirm 7 The next display will show the temperature calibration mode The display should show C CAL 8 If you do not wish to conduct temperature calibration press ENTER to proceed with the ...

Page 4: ...how that auto off function is enabled 13 Press the CAL key to disable this feature and press ENTER to confirm 14 The final setting is for the Reset function The display should show NO RST 15 Press the CAL key to enable or disable the function and press ENTER to confirm Remember that enabling this functions will restore the tester to factory default setting and erase all selected parameters 16 The ...

Page 5: ... calibration point 1 4 Dip the conductivity electrode into the prepared buffer solution 1413 µS cm The end of the sensor must be completely submerged Stir the tester 1 5 Press the ENTER to confirm The tester should recognize the value of the used solution and the displayed value should begin to flash 1 6 If the displayed value is incorrect press the CAL button to set each digit and press ENTER to ...

Page 6: ... the display shows END and completes the calibration process Three points Calibration 3 1 Make sure to select three points calibration in the settings 3 2 Do steps 1 2 to 1 5 and then steps 2 3 to 2 6 3 3 When the second calibration point is completed the display will show CAL3 as the tester prompts you to continue with the third point of calibration 3 4 Rinse the electrode with distilled water an...

Page 7: ...D key again to disable the hold function ELECTRODE SENSOR MAINTENANCE 1 It is recommended to keep the sensitive membrane of the sensor wet This is accomplished by periodically dipping the sensor in distilled water or by using an electrode storage solution when the tester is not in use 2 If the instrument will be kept in storage for a long time place the sensor in an electrode storage solution 3 Ma...

Page 8: ...S cm 84µS cm 1413µS cm 12 88mS cm Conductivity Range 100 2000µS cm 2 00 20 00 20 00 200 0mS cm Accuracy 1 F S Resolution 0 01 0 1 1 Calibration points 1 to 3 points Calibration solutions 10µS cm 84µS cm 1413µS cm 12 88mS cm Temperature Range 0 60 C 32 140 F Accuracy 1 C Resolution 0 1 C 0 1 F Calibration point 1 Calibration range Measured value 10 C Others Temperature compensation 0 60 C 32 140 F ...

Reviews: