8. Troubleshooting
MKC-710 101
For reproducibility of measurement results, moisture content and sample size
relations in below chart is important.
Moisture Content
Sample Size
50
~
100 %
10
~
50 %
1
~
10 %
0.1
~
1 %
0.01 ~
0.1 %
0.001 ~
0.01 %
0.0001
~0.001 %
10
mg
10
~
20 mg
10
~
50 mg
10
~ 100 mg
100mg ~
1.0 g
1
~
10
g
10
~
20
g
Glass contact area jammed
8-6.
If grease on the glass contact areas becomes hard and the respective parts are difficult
to separate, take the following steps;
1)
Discharge anolyte and catholyte.
2)
When using a glass port plug, warm it up with a hair dryer or something similar
to soften KF grease before removing. When using a PTFE port plug, remove it
after cooling the titration flask in a freezer for about five (5) minutes.
Warning!
Do not try and open solidified jointed parts by force.
Glassware may break into piercing pieces for injury.
Do not warm up a PTFE port plug when removing it as doing
so may inflate the material and the titration flask may be
broken.