5. Function Tools
MKC-710 61
5-2-2.Measurement procedure
Measure bromine number and bromine index.
1)
Discharge 20
a
100µL check solution into the electrolysis cell, and press [Pre-titr.]
key.
2)
After pre-titration is finished, press [Sample] key to enter sample name, its ID and
size, and enter the dilution coefficient (D) .
Press [MENU/HOME] to Main display.
3)
Press [Start] key, and discharge the sample into the cell.
Again press [Start] key for titration to start.
4)
After titration is over, the measurement results appear on display with printout when
a printer is connected.
Note
When discharged samples exceed 100mL in total, change the anolyte.
Changing the anolyte may be required each time for those samples, which do
not dissolve in anolyte or measurement results deviate substantially each time.