4. Method
40
[t(stir)]
Select a time length to wait for titration (electrolysis) start after a sample is
discharged into the titration cell, particularly for those samples which are hard to
dissolve in the anolyte and difficult to extract moisture.
0
~
99999s
[t(wait)]
Enter the shortest time length for a measurement from start to end in case a plural
number of inflection points exist or unexpected endpoint is detected due to
fluctuating drift level.
15
~
99999s
[t(max)]
Limit a time length for a titration (electrolysis). This means the total time length
from start to end is [t(stir)] + [t(wait)]+ [t(max)].
The printing covers the measuring process.
[t(max)=0] means titration will not terminate by time limit. This is useful when an
oven is connected and End point is hard to be found due to drift.
0
~
99999s