Chapter 7: Adjust the address of Load Cell
If address needs to be changed, the indicator should turn off first. Connect the load cell
needed to change address and disconnect others.
(
Note
:
the indicator can only connect
one load cell.
),
Plug in the calibration plug, turn on the indicator and begin to operate.
Steps
Operation
Display
Note
1Turn on the
indicator and
choose operating
mode
Press [Amend
Code]
Enter
[1][1][1][1][1][1]
,
Press[Enter]
[-------]
[111111]
Ask to enter password
Only when the password is 111111, can
the next step continue
,
otherwise back
to the weighing state
;
2 Set load cell
address
Press [0][2]
Press [Enter]
[No **]
[No 02]
Display original set load cell serial
number.
Set the address to 2, 01~16 is available
Press [Enter] to confirm to finish the
setting of the address and enter the next
step.
3
.
Display load
cell ISN(weight
value)
Press [Calibration]
[******]
Display load cell ISN
If press [Calibration], then back to the
first step
Power off to finish the modification of
load cell address.
▲
!
Pull out the calibration plug after operation.
▲
!
Error caused by power intermit, please press [check] or shut down
power and reset.
。