www.kelichina.com
Page 19 of 38
Steps
Operation
Display
Note
1.Set division
Value
Press
[Calibration]
Press[0][5]
Press[Enter]
Press[1][0]
[
d1
**]
[d1 05]
[
d2
**]
[d2 10]
The Calibration plug has been plugged in
under weighing state.
Means setting the first part division value;
Set the first part division value to 05, 01~99
available. Press [Check] to skip this step,
press [Weight] back to the weighing state
Press [Enter] to finish the first part division
value setting and enter the second division
value setting;
Set the second part division value to 10,
01~99 available. Press [Check] to skip this
step, press [Weight] back to the weighing
state
2. Set auto zero
tracing
coefficient
Press[2].[0]
Press[Enter]
[Auto *.*]
[Auto 2.0]
Display original set auto zero tracing
coefficient
Set zero tracing to 2.0d, 0.1d~9.9d available.
Press [Check] to skip this step, press [Weight]
back to the weighing state
Press [Enter] to finish the setting of auto zero
tracing coefficient.
3.Set Zero
range
Press[0][1]
Press[Enter]
[CLr **]
[CLr 01]
Display original Zero
Set Zero to 01, 00/01/02/03/04/ available. Start
Zero are separately 4%
、
10%
、
20%
、
40%
、
100%
,
manual 1%
、
2%
、
4%
、
10%
、
20%.
Press [Check] to skip this step, press [Weight]
back to the weighing state
Press [Enter] to finish the setting of Zero
range.