Calibration
Models 2510 and 2510-AT
3.
Select the DMM DC voltage function, and enable auto-range.
4.
Press ENTER. The unit will then prompt for the +8.5V step:
VOLTAGE CAL
Press ENTER to 8.5V
NOTE
The output will turn on automatically when you perform the next step.
5.
Press ENTER. The Model 2510 will 8.5V and simultaneously display the
following:
DMM RDG: +08.50000 V
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
6.
Note and record the DMM reading, and then use the EDIT keys to adjust the
Model 2510 display to agree exactly with the actual DMM reading.
7.
After adjusting the display to agree with the DMM reading, press ENTER. The instru-
ment will then display the following:
VOLTAGE CAL
Press ENTER to output 0.0V
8.
Press ENTER. The Model 2510 will source 0V and at the same time display the
following:
DMM RDG: +00.00000 V
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
9.
Note and record the DMM reading, and then adjust the Model 2510 display to agree
with that reading.
10.
After adjusting the display value to agree with the DMM reading, press ENTER. The
unit will then display the following:
VOLTAGE CAL
Press ENTER to output -8.5V
11.
Press ENTER. The Model 2510 will source -8.5V and display the following:
DMM RDG: -08.50000 V
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
12.
Note and record the DMM reading, adjust the Model 2510 display to agree with the
DMM reading, and press ENTER.
13.
Disconnect the DMM from the OUTPUT terminals.
2510-902-01 (E - Mar 2010)(Service).book Page 10 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...