1-3
Performance Verification
Models 2510 and 2510-AT
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. Typically,
allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature
range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Line power
The Model 2510 requires a line voltage of 100 to 240V and a line frequency of 50 or 60Hz.
Verification tests must be performed within this range.
Recommended test equipment
Table 1-1 summarizes recommended verification equipment and pertinent specifications.
You can use alternate equipment as long as that equipment has specifications at least as good as
those listed in Table 1-1. Keep in mind, however, that test equipment uncertainty will add to the
uncertainty of each measurement. Generally, test equipment uncertainty should be at least four
times better than corresponding Model 2510 specifications.
Table 1-1
Recommended verification equipment
Description
Manufacturer/Model
Specifications
Digital Multimeter
Keithley 2002
DC Voltage
1
20V:
±6.8ppm
Resistance
1
20
Ω
:
200
Ω
:
2k
Ω
:
20k
Ω
200k
Ω
:
±23ppm
±19ppm
±7.4ppm
±7.4ppm
±29.8ppm
Resistors
2
Isotec RUG-Z-1R00-0.1
1
Ω
, ±0.1%, 100W
100
Ω
, ±1%
1k
Ω
, ±1%
10k
Ω
, ±1%
100k
Ω
, ±1%
INPUT/OUTPUT
Mating Connector
3
Keithley CS-846
1. Ninety day, full-range accuracy specification of ranges required for various measurement points.
2. Characterize all resistors using 4-wire ohms function of recommended DMM before use.
3. One connector supplied with Model 2510.
2510-902-01 (E - Mar 2010)(Service).book Page 3 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...