B-6
Calibration Reference
Models 2510 and 2510-AT
SENSe commands
:VOLTage
(:CALibration:PROTected:SENSe:VOLTage)
(:CALibration:PROTected:SENSe:VOLTage:DATA?)
Purpose
To calibrate the Peltier voltage measurement.
Format
:cal:prot:sens:volt <DMM_Reading>
Parameters
<DMM_Reading> =
3.5 to 12.5[V] = cal at + F.S.
-2.5 to 2.5[V] = zero cal
-3.5 to -12.5[V] = cal at - F.S.
Query
:cal:prot:sens:volt:data?
Description
The :CAL:PROT:SENS:VOLT command calibrates the Model 2510 Peltier
voltage measurement. During the calibration process, this command is sent
three times; once each with parameters of approxiF.S, 0, and -F.S.
The actual parameters are voltage readings taken with a precision DMM.
The :CAL:PROT:SENS:VOLT:DATA? query allows you to request the
Peltier voltage calibration parameters.
Example
:CAL:PROT:SENS:VOLT 8.51
Calibrate positive voltage.
:CURRent
(:CALibration:PROTected:SENSe:CURRent)
(:CALibration:PROTected:SENSe:CURRent:DATA?)
Purpose
To calibrate the current measurement.
Format
:cal:prot:sens:curr <Resistor_Value>
Parameters
<Resistor_Value> = 0.5 to 1.5[
Ω
]
Query
:cal:prot:sens:curr:data?
Description
The :CAL:PROT:SENS:CURR command calibrates the Model 2510 cur-
rent measurement. During the calibration process, a 1
Ω
(nominal) resistor
must be connected to the OUTPUT terminals, and the source is automati-
cally set to +4.5V, 0V, and -4.5V. The command parameter is the character-
ized value of the 1
Ω
resistor.
The :CAL:PROT:SENS:CURR:DATA? query allows you request the cur-
rent calibration parameters.
Example
:CAL:PROT:SENS:CURR 1.01
Calibrate current.
2510-902-01 (E - Mar 2010)(Service).book Page 6 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...