Models 2510 and 2510-AT
Calibration Program
C-3
General program instructions
1.
With the power off, connect the Model 2510 and the digital multimeter to the IEEE-488
interface of the computer. Be sure to use shielded IEEE-488 cables for bus connections.
2.
Turn on the computer, the Model 2510, and the digital multimeter. Allow the Model
2510 and the multimeter to warm up for at least one hour before performing calibration.
3.
Make sure the Model 2510 is set for a primary address of 15. (Use the front panel
MENU/COMMUNICATIONS/GPIB selection to check or change the address.)
4.
Make sure the digital multimeter primary address is set to 16.
5.
Make sure that the computer bus driver software (CECHP.EXE) is properly initialized.
6.
Enter the QBasic editor, and type in the program below.
NOTE
Be sure to change the resistor parameters (R$, R$(1) - R$(4)) to the characterized
values of the 1
Ω
, 100
Ω
, 1k
Ω
, 10k
Ω
, and 100k
Ω
resistors. (See Section 2 for details.)
7.
Check thoroughly for errors, then save it using a convenient filename.
8.
Run the program, and follow the prompts on the screen to perform calibration. For test
connections, refer to the following figures in Section 2:
• Voltage calibration DMM connections: Figure 2-1.
• Current and current protection calibration 1
Ω
resistor connections: Figure 2-2.
• Thermistor and RTD calibration 100
Ω
to 100k
Ω
resistor connections: Figure 2-3.
• Shorted INPUT connections: Figure 2-4.
• I-SS and V-SS calibration resistor/DMM connections: Figure 2-5.
• AC ohms calibration 100
Ω
resistor connections: Figure 2-6.
2510-902-01 (E - Mar 2010)(Service).book Page 3 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...