2-24
Calibration
Models 2510 and 2510-AT
4.
With the 1
Ω
resistor still connected, send the following command to calibrate the cur-
rent protection DAC:
:CAL:PROT:IPD
5.
Send this command to turn off the output:
:OUTP OFF
6.
Disconnect the 1
Ω
resistor from the OUTPUT terminals.
Step 4: Temperature calibration
1.
Connect the 100
Ω
resistor to the Model 2510 INPUT terminals, as shown in Figure 2-3.
2.
Send the following command to select the thermistor type temperature sensor:
:SENS:TEMP:TRAN THER
3.
Send the following command to enable 4-wire sensing:
:SYST:RSEN ON
4.
Send this command to select the 100
Ω
resistance range:
:SENS:TEMP:THER:RANG 100
5.
Send the following command with the characterized 100
Ω
resistance value as the
parameter to calibrate the 100
Ω
range:
:CAL:PROT:SENS:TEMP <Resistance_Value>
6.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
7.
Connect the 1k
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
8.
Send this command to select the 1k
Ω
resistance range:
:SENS:TEMP:THER:RANG 1e3
9.
Send the following command with the characterized 1k
Ω
resistance value as the param-
eter to calibrate the 1k
Ω
range:
:CAL:PROT:SENS:TEMP <Resistance_Value>
10.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
11.
Connect the 10k
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
12.
Send this command to select the 10k
Ω
resistance range:
:SENS:TEMP:THER:RANG 1e4
13.
Send the following command with the characterized 10k
Ω
resistance value as the
parameter to calibrate the 10k
Ω
range:
:CAL:PROT:SENS:TEMP <Resistance_Value>
14.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
15.
Connect the 100k
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
2510-902-01 (E - Mar 2010)(Service).book Page 24 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...