Calibrating the Analog Output Subsystem
This section describes how to use the KUSB-3100 Calibration Utility
to calibrate the analog output subsystem of a KUSB-3100 module.
To calibrate the analog output circuitry, you need to connect an
external precision voltmeter to analog output channels 0 and 1 of the
KUSB-3100 module.
Perform the following steps to calibrate the analog output circuitry:
1.
Select the D/A Configuration tab of the KUSB-3100 Calibration
Utility.
2.
Connect an external precision voltmeter to Analog Output 0
(DAC Ch0) of the KUSB-3100 module.
3.
Adjust the offset by entering values between 0 and 63 in the DAC
0 Offset edit box or by clicking the up/down buttons until the
voltmeter reads 0 V.
4.
Connect an external precision voltmeter to Analog Output 1
(DAC Ch1) of the KUSB-3100 module.
5.
Adjust the offset by entering values between 0 and 63 in the
DAC 1 Offset edit box or by clicking the up/down buttons until
the voltmeter reads 0 V.
Note:
At any time, you can click
Restore Factory Settings
to reset
the D/A calibration values to their original factory settings. This
process will undo any D/A calibration settings.
Once you have finished this procedure, the analog output circuitry is
calibrated. To close the KUSB-3100 Calibration Utility, click the close
box in the upper, right corner of the window.
Summary of Contents for KUSB-3100
Page 10: ...Contents x...
Page 14: ...About this Manual xiv...
Page 15: ...1 1 Overview Features 2 Supported Software 3...
Page 18: ...Chapter 1 4...
Page 46: ...Chapter 2 32...
Page 47: ...33 3 Supported Device Driver Capabilities...
Page 86: ...Chapter 4 72...
Page 87: ...73 5 Troubleshooting General Checklist 74 Service and Support 77...
Page 100: ...Chapter 6 86...
Page 101: ...87 A Specifications...
Page 112: ...Appendix A 98...
Page 113: ...99 B Connector Pin Assignments...
Page 115: ...101 C Reading from or Writing to the Digital Registers...