Calibration
83
6
6
6
6
6
6
6
6
6
Note:
At any time, you can click
Restore Factory Settings
to reset
the A/D calibration values to their original factory settings. This
process will undo any auto or manual calibration settings.
Using the Manual Calibration Procedure
If you want to manually calibrate the analog input circuitry instead of
auto-calibrating it, perform the following steps:
1.
Adjust the offset by performing the following steps:
a.
Verify that 0V is applied to AD Ch0, and that A/D Channel
Select is set to Channel 0.
The current voltage reading for this channel is displayed in the A/D
Value window.
b.
Adjust the offset by entering values between 0 and 63 in the
Offset edit box, or by clicking the up/down buttons until the
A/D Value is 0 V.
2.
Adjust the gain by performing the following steps:
a.
Verify that 9.375V is applied to AD Ch0, and that A/D
Channel Select is set to Channel 0.
The current voltage reading for this channel is displayed in the A/D
Value window.
b.
Adjust the gain by entering values between 0 and 63 in the
Gain edit box, or by clicking the up/down buttons until the
A/D Value is 9.3750.
Summary of Contents for KUSB-3100
Page 10: ...Contents x...
Page 14: ...About this Manual xiv...
Page 15: ...1 1 Overview Features 2 Supported Software 3...
Page 18: ...Chapter 1 4...
Page 46: ...Chapter 2 32...
Page 47: ...33 3 Supported Device Driver Capabilities...
Page 86: ...Chapter 4 72...
Page 87: ...73 5 Troubleshooting General Checklist 74 Service and Support 77...
Page 100: ...Chapter 6 86...
Page 101: ...87 A Specifications...
Page 112: ...Appendix A 98...
Page 113: ...99 B Connector Pin Assignments...
Page 115: ...101 C Reading from or Writing to the Digital Registers...