![Keithley KUSB-3100 User Manual Download Page 96](http://html1.mh-extra.com/html/keithley/kusb-3100/kusb-3100_user-manual_3961323096.webp)
Chapter 6
82
Calibrating the Analog Input Subsystem
This section describes how to use the KUSB-3100 Calibration Utility
to calibrate the analog input subsystem of a KUSB-3100 module.
Connecting a Precision Voltage Source
To calibrate the analog input circuitry, you need to connect an
external precision voltage source to Analog In 0 (AD Ch0) of the
KUSB-3100 module.
Using the Auto-Calibration Procedure
Auto-calibration is the easiest to use and is the recommended
calibration method. To auto-calibrate the analog input subsystem,
perform the following steps:
1.
Select the
A/D Configuration
tab of the KUSB-3100 Calibration
Utility.
2.
Set the voltage supply on AD Ch0 to 0 V.
3.
Click
Start Auto Calibration
.
A message appears notifying you to verify that 0 V is applied to AD
Ch0.
4.
Check that the supplied voltage to AD Ch0 is 0V, then click
OK
.
The offset value is calibrated. When the offset calibration is complete, a
message appears notifying you to set the input voltage of AD Ch 0 to
+9.375 V.
5.
Check that the supplied voltage to AD Ch0 is +9.375V, then click
OK
.
The gain value is calibrated.
6.
Click OK to finalize the analog input calibration process.
Summary of Contents for KUSB-3100
Page 10: ...Contents x...
Page 14: ...About this Manual xiv...
Page 15: ...1 1 Overview Features 2 Supported Software 3...
Page 18: ...Chapter 1 4...
Page 46: ...Chapter 2 32...
Page 47: ...33 3 Supported Device Driver Capabilities...
Page 86: ...Chapter 4 72...
Page 87: ...73 5 Troubleshooting General Checklist 74 Service and Support 77...
Page 100: ...Chapter 6 86...
Page 101: ...87 A Specifications...
Page 112: ...Appendix A 98...
Page 113: ...99 B Connector Pin Assignments...
Page 115: ...101 C Reading from or Writing to the Digital Registers...