Section 4: TSP command reference
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
4-2
DMM6500-905-01 Rev. E June 2021
cal.adjust.count
This attribute returns the number of times the instrument has been adjusted.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Nonvolatile memory
Not applicable
Usage
adjustments
= cal.adjust.count
adjustments
The number of adjustments
Details
You can use this command if calibration is locked or unlocked.
The adjust count is read-only. The count is automatically incremented by one when the
cal.save()
command is sent with calibration unlocked.
Example
Count = cal.adjust.count
print(Count)
Assign the number of times the instrument has been adjusted
to a user variable named
Count
.
Output the value. Example output:
3
This shows that the instrument has been adjusted 3 times.
Also see
cal.adjust.date
(on page 4-3)
Summary of Contents for DMM6500
Page 1: ...6 Digit Multimeter with Scanning Calibration and Adjustment Manual DMM6500...
Page 5: ......
Page 6: ......
Page 64: ......