Section 2: Performance verification
DMM6500-905-01 Rev. E June 2021
2-49
Digitize current 10 A range verification
Ensure cabling is designed to conduct 10 A without significant voltage drop. In general,
cables for the 10 A range steps should be made of heavy gauge wire and should be as short
as practical.
To verify the 10 A range:
1. Connect DMM6500, calibrator, and amplifier as shown in the following figure.
Figure 19: Connections for the 10 A range