Section 3: Adjustment
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
3-24
DMM6500-905-01 Rev. E June 2021
10 mA range (-) full scale nominal
cal_DCI_10mA_fs_neg
dc current 0.022
-0.01
8508A or 8588A
Reading
10 mA
cal_DCI_DIGI_10mA_fs_neg
dc current 0.022
-0.01
8508A or 8588A
Reading
10 mA
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters
Calibrator
function
Calibrator range Calibrator
value
Calibrator
stimulus
value
Reference
DMM dc
current range
cal_DCI_100mA_fs_pos
dc current 0.22
0.1
n/a
20 A
cal_DCI_DIGI_100mA_fs_po
s
dc current 0.22
0.1
n/a
20 A
cal_DCI_100mA_fs_neg
dc current 0.22
-0.1
n/a
20 A
cal_DCI_DIGI_100mA_fs_ne
g
dc current 0.22
-0.1
n/a
20 A
cal_TS5
dc current n/a
n/a
n/a
20 A
cal_dcI_1A_fs_pos
dc current 2.2
1
n/a
20 A
cal_DCI_DIGI_1A_fs_pos
dc current 2.2
1
n/a
20 A
cal_DCI_1A_fs_neg
dc current 2.2
-1
n/a
20 A
cal_DCI_DIGI_1A_fs_neg
dc current 2.2
-1
n/a
20 A
cal_DCI_DIGI_3A_fs_pos
dc current 2.2
2
n/a
20 A
cal_DCI_DIGI_3A_fs_neg
dc current 2.2
-2
n/a
20 A
DC current 10 A range adjustment
The following procedure provides instructions for completing a rear-terminal dc current adjustment for
the 10 A range. The following section provides a command parameter table to complete the
adjustment.
For this adjustment, you need a:
•
Fluke 5720 calibrator
•
10 A cable