DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 2: Performance verification
DMM6500-905-01 Rev. E June 2021
2-37
Verify digitize current 1 A range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
Full scale (+)
1
0.999
1.001
Half scale (+)
0.5
0.49935
0.50065
Half scale (–)
–0.5
-0.50065
-0.49935
Full scale (–)
–1
-1.001
-0.999
Verify digitize current 3 A range
Description
Calibrator setpoint
(A)
Lower limit
Upper limit
*Full scale (+)
2
1.997
2.003
Half scale (+)
1.5
1.49745
1.50255
Half scale (–)
–1.5
–1.50255
–1.49745
*Full scale (–)
–2
–2.003
–1.997
* The 3 A range full-scale test points are limited to 2 A in this table because of the accuracy
limitations of Fluke Models 57xxA and 5725A series calibrators at currents above 2.2 A.
Verify digitize current 10 A range
See
Digitize current 10 A range verification
(on page 2-49) under
Rear-panel verification
(on page
2-46).
AC current verification
The following topics describe how to verify ac current.